Accurate, Reliable Sodium Monitoring
Combination reference and pH electrode confirms reagent addition for reliable operation, minimizes reagent consumption, and reduces maintenance.
Configurable, Automated Calibration
Unattended, automated calibration ensures the accurate measurement of sodium in your water system and detects breakthroughs right when they occur.
Open Panel Design Reduces Costs
The open panel design reduces costs in areas where additional protection is not required, allowing customers to easily meet budget constraints.
Material No.: 30768009
Sodium Measurement Range | 0.001 - 100,000 ppb or equivalent ppm, autoranging |
Sodium Measurement Accuracy | ±10% of reading or ± 0.05 ppb whichever is greater |
Response Time | 90% at 5 minutes |
Sample pH | 2.5-12 |
Update Rate | Once per second |
Reagent Consumption | Diisopropylamine (DIPA), approximately 1 L per 3 months more at higher temperatures and for cation exchange samples |
Sample Flow rate | >40 mL/min (>20 mL/min for cation exchange samples), excess to drain |
Sample Temperature | 5-50°C (41-122°F) |
Sample Pressure | 0.3-7 bar (5-100 psig) |
Grab Sample Measurement | Included |
Power Supply | 100-240 VAC, 50-60 Hz, 25 W, on power loss all settings are retained without batteries |
Ambient Operating Temperature | 10-45°C (50-113°F) |
Humidity | 10-90% non-condensing |
Process Connection | Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included Sample Inlet: 1/4" or 6 mm OD tube SS compressin fitting Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included |
Operator Interface | 4-line backlit LCD, 5 tactile keys simultaneous display of sodium, conditioned pH, auto-cal status (temperature optional) |
Analog Output | 22mA alarm, 500 ohm max load, not for use with externally powered circuit For sodium, conditioned pH, temperature four powered 0/4-20 mA, |
Analog Output Scaling | Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
Measurement Parameter | Sodium |
Electrode Conditioning | Part of auto-cal sequence |
Relay Contacts | 3A resistive freely assignable to setpoints for sodium, pH, temperature other relays used for auto-cal |
Sodium Calibration | Automatic, unattended 3-point known addition |
Calibration | Automatic |
Analog Output Accuracy | ±0.05 mA |
Model Family | 2300Na |
EnclosureType | Panel |
Built-in sequencer | No |