Researchers in life sciences and (bio)chemistry constantly pursue methods that enhance efficiency and precision while delivering accurate results and streamlining complex workflows. A prime example is the automated determination of the acid dissociation constant (pKa), which illustrates how contemporary technology can improve traditional techniques. The strength of this automated approach is its capacity to provide quick and dependable results without sacrificing accuracy. This article highlights the user-friendly aspects of UV/Vis-titration workflows, which reduce the need for manual intervention and decrease reagent consumption.
![]() |
The acid dissociation constant (pKa) is an essential parameter in the life sciences, particularly within the pharmaceutical and agrochemical sectors. It measures the strength of an acid in a given solvent, reflecting its tendency to release hydrogen ions (H⁺) into the solution. A solid understanding of pKa values is vital for characterizing active pharmaceutical ingredients (APIs), as their ionization states significantly impact absorption and effectiveness in biological systems. Traditional techniques for determining the pKa value include potentiometric titration, conductometry, spectrophotometry, nuclear magnetic resonance (NMR), and calorimetry. Potentiometric titration is the most popular due to its simplicity and rapid results, involving the straightforward addition of a titrant while monitoring potential changes. In contrast, methods like NMR and calorimetry, though informative, often necessitate more complex setups and longer measurement times, making them less practical than potentiometric titration.
Recent advancements in analytical methods have led to automated workflows that improve pKa value determination by combining titration with UV/Vis spectrophotometry. This integrated approach streamlines the process and reduces reagent consumption and sample waste. An illustrative example is the automated determination of the pKa value for bromothymol blue, a compound chosen for its distinct color change across a broad pH range, facilitating the visualization of the transition between its protonated and deprotonated forms. Researchers calculate the pKa value from the inflection point in the absorbance versus pH plot, which they obtain by measuring the molecule's absorption spectrum at different pH values. This method is a reliable alternative to traditional techniques, delivering consistent results while minimizing the workload for operators.
Experimental Setup Made Simple
The methodology employs a fully automated system controlled by LabX software, ensuring a user-friendly experience from sample preparation to UV/Vis measurement and titration.
The system integrates the Titration Excellence T9 with the UV5 Excellence Spectrophotometer. The DispenSix Liquid Handler manages the dispensing and aspiration of solutions, including KH2PO4 and Na2HPO4 buffer solutions, as well as water for rinsing and waste removal. The InMotion Pro Autosampler automates sample handling with dedicated beakers for conditioning and rinsing, equipped with multiple pumps (SP280, SD660, SPR200) for accurate liquid transfer. Additional resources include sensors and various titrants, such as H2O, bromothymol blue (BTB), NaOH, and HCl, for the analysis.
The system automatically prepares the bromothymol blue solution during the sample preparation phase. Once this preparation is complete, measurements are conducted on both devices using a synchronized method.
The sample's pH value is measured directly in the titration beaker on the autosampler. Then, the SPR200 pump transfers the sample into the flow cell for UV/Vis spectrum measurement. After the measurement is complete, the solution is returned to the sample beaker.
![]() |
Figure 1: Analytical system for fully automating determination of the pKa value by coupling different instruments. The LabX laboratory software controls the whole system. |
What the Data Tells Us
The study on bromothymol blue utilized absorbance measurements at two distinct wavelengths, λ1 = 432 nm and λ2 = 617 nm, corresponding to the highest absorbance values at low and high pH levels. Each wavelength reflects the distinct absorbance characteristics of the indicator's protonated and deprotonated forms.
The study demonstrated key findings regarding the absorbance measurements of bromothymol blue (BTB) across different pH values. BTB was chosen for this analysis because its well-known pKa value makes it an ideal candidate for validating the methodology used in the research. As shown in Figure 2, at higher pH values, the absorbance peaks at 617 nm indicate that the deprotonated form of BTB is predominant in alkaline conditions. Conversely, at lower pH values, the peak absorbance occurs around 432 nm, where the protonated form is more prevalent.
Through linear regression analysis of absorbance data against pH values, the average pKa value obtained was 7.20 ± 0.03. This result aligns well with established literature values ranging from 6.8 to 7.5, confirming the accuracy and reliability of the method employed. The linear relationship between absorbance and pH value is illustrated in Figure 3, further supporting the findings and demonstrating the effectiveness of the dual-wavelength strategy for evaluating pKa in BTB solutions.
![]() |
Figure 2: UV/Vis spectra of aqueous solutions of BTB at 12 different pH values. |
![]() |
Figure 3: Calculation of the pKa value of BTB by linear regression. The pKa value is given by the intercept of the line with the y-axis (pH), i.e., where pKa = pH. |
The Smart Way to Measure pKa Value: Automation in Action
METTLER TOLEDO proudly introduces a fully automated analytical platform for precise pKa value determination. This innovative system features advanced instruments, such as the Titration Excellence and UV/VIS Excellence Spectrophotometer, along with equipment like the DispenSix Liquid Handler and InMotion Autosampler, all seamlessly controlled by LabX software.
By combining these technologies, the platform significantly boosts operational efficiency and data reliability while reducing the potential for human error. The platform specifically addresses the needs of modern laboratories by automating both titration and spectroscopic analysis, which allows researchers to characterize molecular behavior more accurately across various chemical environments.
In summary, this state-of-the-art automated platform streamlines workflows and empowers researchers to achieve exceptional results in acid dissociation studies, paving the way for groundbreaking discoveries in analytical chemistry.
![]() |
Dr. Ashish Nadar |
Dr. Ashish Nadar is Deputy Manager of Content Writing at IMSG-SBU AnaChem, where he leverages his expertise in chemistry to craft impactful scientific content.
With a PhD in Chemistry from BARC, India, and postdoctoral research experience at IIT Bombay and TIFR Mumbai, Dr. Nadar brings a robust academic foundation to his role. Over the past three years, he has contributed to AnaChem’s success, initially combining his skills as an Application Specialist and Content Writer before transitioning to his current thought leadership position.