Tiếng Việt
OptiMax Starter Kit and HFCal

OptiMax Starter Kit and HFCal

Material No.: 511628301

Gọi để được báo giá

Thông số kỹ thuật

Các thông số - OptiMax Starter Kit and HFCal

Trọng lượng
35 kg
Thể tích Vận hành
150 mL – 1.000 mL
Kích thước (CxRxS)
78 cm x 39 cm x 42 cm
Giới hạn nhiệt độ
-40 °C – 180 °C
Chiếu đèn Phía sau / Phía Trước
Yes/Yes
Precision Heat Transfer
Typically +/- 3%
Accuracy and Precision Specific Heat
Typically +/-10%
Accuracy Heat Flow
Based on comparison of qr_hf with qc resp. ∫qr_hf with ∫ qc
Isothermal conditions: ± 3 % to 5 %
Non-isothermal conditions: ± 5 % to 10%
Calorimetry Type
Heat Flow
Vật liệu
OptiMax Starter Kit (51162830) and Upgrade Kit OptiMax HFCal (30050150)
Chế độ Nhiệt độ
Jacket control, reaction mixture control, distillation control, crystallization control
Gia nhiệt / Làm lạnh
Bộ ổn nhiệt Trạng thái Rắn
Yêu cầu Nguồn điện
100 – 240 VAC, 50/60Hz

Tài liệu liên quan

Bảng dữ liệu

EasyMax (102, 102 LT, 402) HFCal and OptiMax HFCal Datasheet
Datasheets for EasyMax and OptiMax HFCal models include technical specifications and configuration options.

Tài liệu quảng cáo sản phẩm

Chemical Synthesis Reactors Product Catalog
View the full catalog of chemical synthesis reactors, workstations, and accessories. Find what you need for your lab!

Thông tin sản phẩm mở rộng

OptiMax Operating Instructions
Guide on how to operate your OptiMax chemical synthesis workstation.
OptiMax Service
METTLER TOLEDO provides a range of value-added service options to help ensure instrument uptime, provide quality results, and ensure regulatory compli...

Phụ kiện

Vật tư tiêu hao

Phần mềm

More Information

Thermostat 
HeatingElectrical
CoolingPeltier
Cooling Performance0°C   35 W 
20°C 65 W
40°C   95 W
Temperature Range-40°C to 180°C (jacket temperature)
Control Modesisothermal and isoperibolic, constant or ramp, reflux, distillation & crystallization
Stirring30 rpm to 1200 rpm
DimensionsThermostat: 388 mm x 414 mm x 539 mm
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz, 1300 VA
OperationiControl software for calorimetric applications
Graphical touchscreen and/or iControl Software
 
  
Reactors 
TypeSingle piece or two piece (with bottom drain valve)
Volume1000 mL
Pressure50 mbar to 1 bar
MaterialDuran® glass
StirrerPitch-blade (glass, Alloy 22)
Anchor (glass, Alloy 22)
Half-moon (PTFE for single piece reactor)
  
Calorimetry 
TypeHeat Flow
Calibration HeaterMax. 20 W
Precision Heat TransferTypically ± 3%
Accuracy and Precision Specific HeatTypically ± 10%
Accuracy Heat FlowTypically ± 3% to 5%, under isothermal conditions
Typically ± 5% to 10%, under non-isothermal conditions
Sensitivity≤0.1 W measured as standard deviation of qr under steady state conditions.
Peak-to-peak noise is ≤ approximately 0.4 W.
  
HFCal Module 
Power and Data InterfaceDirect connection to CAN interface of standard OptiMax,
no additional power supply or PC connection required
Data LoggingVia iControl Software
Dimensions120 mm x 40 mm x 170 mm
  
SoftwareiControl 5.2 or higher required, HFCal license
Nhận Báo giá
OptiMax Starter Kit and HFCal