Withstands Demanding Applications
The InPro 4262i uses a polymer reference system and HF resistant glass for precise pH measurement, even under difficult industrial conditions.
Advanced Predictive Diagnostics
This pH sensor has Intelligent Sensor Management (ISM) technology, offering predictive diagnostics to reduce maintenance and simplify calibration.
Accurate pH & ORP Measurement
The InPro 4262i provides a measurement range from 1-14 pH and has a solution ground for redox measurement and enhanced stability.
Material No.: 30018467
The InPro 4262i features an open junction Argenthal reference system with a Xerolyt® Extra polymer electrolyte that eliminates sensor clogging and ensures the sensor is resistant to poisoning, solvents, and strongly acidic or basic conditions.
This pH sensor features intelligent diagnostics to help manage your maintenance. The sensor provides a clear display of its remaining lifetime, time remaining until calibration is needed and indicates when the sensor will need to be cleaned.
The InPro 4262i is ATEX and FM certified for hazardous areas. It withstands temperatures up to 176°F (80°C) and uses a special hydrofluoric, acid resistant glass membrane.
Ensure the safety of the maintenance team with the ability to Plug & Measure. The InPro 4262i can be calibrated away from the process and then quickly swapped into the process environment without the need to recalibrate.
ISM (Digital) or Analog | ISM (Digital) |
pH Sensing Technology | Hydrofluoric Acid Resistant Glass Membrane (HF) |
pH Range | 1 – 11 |
Reference System | Xerolyt Extra polymer electrolyte, open double junction, Argenthal cartridge with silver-ion trap |
Operating Temperature (Fahrenheit) | 32 °F - 176 °F |
Operating Temperature (Celsius) | 0 °C - 80 °C |
Cable Connector | K8S |
Process Connection | Pg 13.5 Thread |
Sterilizable/Autoclavable | No |
Sensor Length | 120 mm |
Certificates & Approvals | ATEX/IECEx: Ex ia IIC T6/T5/T4/T3 Ga/Gb CE FM: IS Cl. I, II, III, Div 1, GR ABCDEFG/T6 METTLER TOLEDO Quality Certificate PED |
Operating Pressure (Bar) | 0 to 15 barg @ 25°C 0 to 7 barg @ 130°C |
Model Family | InPro4260(i) |
Temperature Sensor | Digital |
Measurement Parameter | ORP/Redox pH |
Shaft Material | Glass |
Operating Pressure (PSI) | 0 to 101 psig @ 266°F 0 to 217 psig @ 77°F |
Solution Ground/Redox Metal | Platinum |
Refillable Electrolyte | No |
Hazardous Area Approvals | Yes |
Primary Media/Application | Chemical Processes |
Hygienic/Biocompatibility Approvals | No |
The DS AK9-RJ12 is a USB cable adaptor for connecting ISM 1-wire sensors to iSense software. This allows pH, DO and CO2 sensors to be evaluated, maintained and calibrated in the workshop or laboratory.
Diagnostics Away From the Transmitter
The iLink module allows you to access all of the diagnostics of your ISM sensors away from the process with iSense desktop software.
Application Profiling Drives Performance
Store application data on iSense and load directly into an ISM sensor. This allows sensors to learn from preceding sensors in the same process.
Easy Access to Paper Free Documentation
The iLink cable allows sensor calibration and adjustment protocols to be generated automatically in iSense and stored in a fully digital database.
Material No.: 52300383
Material No.: 30130631
Material No.: 51319010
Material No.: 51319011
Material No.: 51340058
Material No.: 51340228
Material No.: 51340060
Material No.: 51340229
Material No.: 51300194
Material No.: 51340230
Material No.: 51340231
Material No.: 51340232
Material No.: 51319012
Material No.: 51319016
Material No.: 51319017
Material No.: 51340055
Material No.: 51340057
Material No.: 51340059
Material No.: 51300193
Material No.: 51340056
pH Sensor InPro4262i/SG/120