The ReactIR DS Instrument Performance Assurance (IPA) module enables users to regularly validate wavenumber accuracy and calibrate, if necessary, to a Polystyrene standard certified by the National Institute of Standards and Technology (NIST). As a result, scientists can be confident that their instruments are in proper working condition, ready when needed and the result can be trusted.
- Simplicity - Plug-n-Play DS interface enables a seamless connection to the ReactIR and removes the need for alignment. Built-in thumbscrews secure the module and the sliding cover eliminates the risk of contamination to the internal Polystyrene standard when not in use.
- Plug-In and Calibrate - Remove existing sampling technology to attach the DS IPA module and immediately begin wavenumber validation and/or calibration. Peak height counts are easily adjusted using an integrated Iris adjustment.
- NIST Traceability - To meet internal and external compliance standards, wavenumber calibration is performed using a NIST traceable 1.5 mil Polystyrene standard with a certification of traceability to 6 NIST 192lb frequencies.
- iC IR Software - ReactIR coupled with iC IR software and the DSA IPA module provides a simple solution for wavenumber calibration. The wizard-based validation and calibration feature guides users through the entire process. Easily archive results for audits with built-in reports generated for both validation and calibration in a non-editable PDF format.