Mastered in no time
Large color touchscreen provides direct measurement and easy data entry. Instructions guide users through the sensor calibration process.
Tracks your records
The instrument stores data like timestamp, sample and sensor ID with every measurement and prints or exports records automatically.
Tough enough
Closed connector compartment and replaceable cover keep the instrument safe from dust and spills, even with sensors attached (IP54 in-use).
Material No.: 30671568
Parameter | Conductivity pH ORP |
Channel | Dual-channel |
Version kit | Ready-to-measure kit |
Sensor | InLab® 741-ISM InLab Pure Pro-ISM |
pH resolution | 0.1 0.01 0.001 |
pH accuracy (±) | 0.002 |
Conductivity measuring range | 0.001 μS/cm – 2000 mS/cm |
Conductivity resolution | 0.001 – 2 |
Conductivity accuracy (±) | 0.5 % |
mV resolution | 1 0.1 |
mV accuracy (±) | 0.1 |
Temperature Range | -30 °C – 130 °C |
Temperature resolution | 0.1 °C |
Temperature accuracy (±) | ±0.1 °C between 0 °C and 100 °C |
Ease of Use | Touch screen Calibration Support Guided Verification Simple Methods Multilingual Menu |
Security and compliance | GLP User management |
Memory size | 2000 standard measurements |
Data Transfer | PC Printer USB stick |
Flexibilty and modularity | Stirrer Barcode reader |
Protection Rating | Watertight and dustproof (IP54) |
Sensor Included | Sensor for pure water |
Display Type | 7" Color TFT |
Meter Type | Benchtop Dual-Channel |
mV measuring range | -2000..2000 mV |
Dimensions | 195x205x65 mm |
Weight | 850 g |
Connectors | BNC RCA / Cinch Mini-DIN (conductivity) USB A (x2) USB B Mini-DIN (for stirrer) |
Compatible with Intelligent sensor management (ISM) | Yes |
Portable Meter | No |
Product Line | SevenDirect |
Pure water specialist sensors
InLab Pure-Pro ISM and InLab 741-ISM are specialist sensors for pH, temperature and conductivity measurements in pure water. They are specifically designed to achieve high accuracy in this challenging sample.
Electrode arm for efficient operation
The EasyPlace electrode arm makes placing the sensor in the sample or buffer container easy. Dedicated positions for cleaning and storage support correct sensor handling. Vertical and horizontal movements facilitate precise positioning, even with one hand.
Protect critical settings
A two-level user management protects the most important settings. Operators retain the flexibility to individually set preferences like language and screen brightness.
Display adapts to your needs
View detailed measurement information on one measurement screen – or focus on the result only and maximize readability with the uFocus mode.
Two Parameters in One Device
Measure pH and conductivity with one instrument – simultaneously or separately as the application requires.
Material No.: 30671569
Material No.: 30671570
Material No.: 30671580
Material No.: 30671583
Material No.: 30671529
Material No.: 30014092
Material No.: 30014094
Material No.: 51344055
Material No.: 30248830
Material No.: 51344072
Material No.: 30014096
Material No.: 51344163
Material No.: 51343151
Material No.: 51344172
Material No.: 30095314
Material No.: 30095315