DO Sensor InPro6860i/12/120/mA/HD

DO Sensor InPro6860i/12/120/mA/HD

Anti-Bubble Dissolved Oxygen Sensor.The InPro 6860i/12/120/mA/HD is an optical dissolved oxygen sensor with a mechanically modified hydrophilic surface. Combined with its tilted design, it prevents bubble accumulation, creating a highly stable measurement signal.

Increased Yield and Product Quality

The angled tip and hydrophilic surface prevent bubble formation. This eliminates noise for improved production efficiency and process control.

For Digital & Analog Bioreactors

This anti-bubble DO sensor has digital (MODBUS, RS-485) and analog (mA) output signals for compatibility with benchtop bioreactors and transmitters.

Sterilizable and Autoclavable

The InPro 6860i/12/120/mA/HD provides fast, precise oxygen measurement, even after repeated autoclaving and sterilization cycles up to 140° C.

Material No.: 30449705

Позвонить специалисту

Характеристики

Stable, Reproducible Yields

Achieving consistent batches and product quality requires reliable oxygen measurement. The hydrophilic surface finish and tilted tip of the oxygen sensor prevent measurement interference from bubble accumulation that can effect yield and product quality.

Hygienic and Durable Design

Avoiding biofouling and ensure a stable signal with the PTFE coating of the OptoCap™ oxygen-sensing element. This ensures that the InPro6860i/12/120/mA/HD has the durability to withstand multiple autoclaving and standard CIP / SIP cycles.

Uncompromised Reliability

This optical DO sensor offers the benefits and predictive diagnostics of Intelligent Sensor Management. Achieve faster start up with Plug and Measure technology. Increased system availability is provided by maintaining a stock of pre-calibrated sensors.

Exceptional Performance

The 120mm InPro6860i/12/120/mA/HD reduces both handling time and running costs due to lower calibration and maintenance needs. This is performance is achieved by an electrolyte-free, one-piece replacement part that is monitored by predictive diagnostics.

Спецификации

Технические характеристики - DO Sensor InPro6860i/12/120/mA/HD

Измеряемый параметр
Кислород
Принцип измерения
Гашение флуоресценции
Погрешность измерения
± (1% от измеренного значения + 8 ppb)
Время отклика (2)
T98= <90 с
Основная среда / область применения
Фармацевтика/биотехнологии
Цифровой (ISM) или аналоговый
Цифровой (ISM)
Противопузырьковая конструкция
Да
Материал металлических смачиваемых частей
Нержавеющая сталь 316L
Качество обработки поверхности смачиваемых частей
N5/Ra16 (Ra=0.4μm/16μin)
Материал уплотнительного кольца
EPDM
Электропитание
24В пост. тока
0.1 A
Подключение к технологической линии
Pg 13.5
Длина
120 mm
Материал смачиваемой мембраны
ПТФЭ
Испольнение коннектора
прямой
Семейство моделей
InPro 6860i
Автоклавируемый/стерилизуемый
Да
Устойчивый к CIP
Да
Механическое отображение измерения
-4 °F
284 °F
Рабочее давление, фунт/кв. дюйм
0-87 PSI
Интеграция с аналоговыми схемами
Интеграция с цифровыми схемами
RS485 ISM® и MODBUS RTU
Кабельный разъем
VP6 / VP8 (аналоговый /цифровой)
Сертификаты соответствия гигиеническим нормам / биологической совместимости
FDA/USP Class 6
Сертификаты для использования во взрывоопасных зонах
ATEX/IECEx: II 1/2G Ex ia/ib HC T6 Ga/Gb II1/2D Ex ia/ib nie T83 °C Da/Db
Сертификаты и разрешения на применение
Quality certificate, Material certificate 3.1, Surface finish certificate, ATEX/IECEx certificate, FDA/USP Class IV
Диапазон измерения
0 - 60% насыщения О2
Стерилизуемый, автоклавируемый
Да

Watch the video on the Anti-Bubble OptoCap

Features and Benefits


Stable, Reproducible Yields
Achieving consistent batches and product quality requires reliable oxygen measurement. The hydrophilic surface finish and tilted tip of the oxygen sensor prevent measurement interference from bubble accumulation that can effect yield and product quality.
 

Hygienic and Durable Design
Avoiding biofouling and ensure a stable signal with the PTFE coating of the OptoCap™ oxygen-sensing element. This ensures that the InPro6860i/12/120/mA/HD has the durability to withstand multiple autoclaving and standard CIP / SIP cycles.
 

Uncompromised Reliability
This optical DO sensor offers the benefits and predictive diagnostics of Intelligent Sensor Management. Achieve faster start up with Plug and Measure technology. Increased system availability is provided by maintaining a stock of pre-calibrated sensors.

Exceptional Performance
The 120mm InPro6860i/12/120/mA/HD reduces both handling time and running costs due to lower calibration and maintenance needs. This performance is achieved by an electrolyte-free, one-piece replacement part that is monitored by predictive diagnostics.
 

Документация

Буклеты об оборудовании

Intelligent Optical Oxygen Sensor Easy Handling – Exceptional Performance
Continuous dissolved oxygen (DO) measurement is critical during fermentation and cell culture. The InPro 6860 i HD Anti-Bubble™ Oxygen Sensor is mecha...

Технические характеристики

Manuals

Operating Instructions: Optical O₂ Simulator
METTLER TOLEDO optical O2 sensors are intended solely for inline measurement of the oxygen partial pressure, as described in this instruction manual.

White Papers

Снижение коррозии
В этом информационном документе описан практический опыт борьбы с коррозией в химической промышленности. В нем раскрыта важная роль измерительных приб...

Примеры применения

Reduce maintenance case study
This case study describes how Ajinomoto needed more dependable pH and DO sensors, plus less sensor maintenance, they turned to METTLER TOLEDO. Now, in...
История успеха: «Компания Novozymes выходит в свет»
Мартин Фальк (Martin Falk) очень доволен эксплуатационными характеристиками датчика InPro 6860 i: «Новая система более стабильна и надежна. В будущем...

Аксессуары

Интеграция цифровых датчиков

M100 Sensor Mount (SM) — одноканальный трансмиттер для подключения к биоконтроллеру оптических ISM-датчиков растворенного кислорода. Интерфейс Bluetooth 4.0 совместим с ПО iSense для ПК и мобильных устройств.

Material No.: 30365367

Средства калибровки и технического обслуживания

This is a USB iSense adaptor for connecting ISM RS485 VP sensors to iSense for pH, DO and CO2 measurement and sensor calibration.

Material No.: 30014134

A universal device for connecting both 1-wire and RS485 ISM sensors to a computer running iSense software.

Material No.: 30130631

Material No.: 52300400

Расходные части

Программное обеспечение

More Information

Оформление запроса
DO Sensor InPro6860i/12/120/mA/HD
FDA