High Operational Availability
The InPro 6960i offers outstanding measurement performance with a low detection limit of 0 ppb to 25 ppm, minimum drift and a short response time.
Suitable for Hygienic Processes
The optical technology is available for various brewery applications. It is especially suitable for sterile and hygienic processes in wort aeration.
Advanced Predictive Diagnostics
This DO sensor has Intelligent Sensor Management (ISM) technology, offering predictive diagnostics to reduce maintenance and simplify calibration.
Material No.: 52206500
With the fast OptoCap replacement and the Plug & Measure ability, the InPro6960i optical oxygen sensor can be calibrated away from the process, and then quickly swapped into the process environment without the need to recalibrate. No polarization is requi
The materials of construction of the InPro6960i DO sensor are FDA compliant and the easy-to-clean high-polished surface (N5 grade) also satisfies increasing regulatory requirements.
The sensor's measurement performance helps to reduce out of spec production. ISM helps to make optical oxygen measurement simple and more reliable and significantly reduces the risk of sensor failure during production.
The built-in digital ISM technology provides the user important information about the sensor wear. Maintenance planning becomes easy, as accurate diagnostics predict when sensor maintenance will be required.
Measurement Parameter | Oxygen |
Measurement Technology | Optical fluorescence quenching |
Accuracy | ± (1 % of the reading + 8 ppb) |
Response Time | T98 = < 20 s |
Primary Media/Application | Food and Beverage |
ISM (Digital) or Analog | ISM (Digital) |
Wetted Metallic Material | 316L stainless steel |
Surface Roughnes of Wetted Material | N5/Ra16 (Ra=0.4μm/16μin) |
Sensor Diameter | 12mm |
O-Ring Material | EPDM |
Process Connection | Pg 13.5 |
Sensor Length | 120 mm |
Wetted membrane material | Silicone |
Connector Design | Straight |
Model Family | InPro 6900i optical series |
Autoclavable/Sterilizable | Yes |
CIP resistant | Yes |
Mechanical Temp. Range | 23 °F – 250 °F (-5 °C – 121 °C) |
Operating Pressure (Bar) | 0.2–12 Bar |
Operating Pressure (PSI) | 02.9–174 PSI |
Design Pressure | 174 psi (12 bar) |
Digital Integration | RS485 - ISM® |
Cable Connector | M12 5-Pin |
Hygienic/Biocompatibility Approvals | FDA/USP Class 7 |
Certificates & Approvals | Quality certificate, Material certificate 3.1, Surface finish certificate, FDA/USP Class IV |
Measurement Range | 0- 25 ppm saturation |
Residual signal in O2 free media | <0.025 of signal in ambient air |
Operating Range | 8 ppb - 60% O2 sat. |
Sterilizable/Autoclavable | Sterilizable |
Measured Pressure Resistance | 3 bar – 12 bar |
Measuring Temp. Range | 23 °F – 104 °F (-5 °C – 40 °C) |
Detection limit | 8 ppb |
Connection | 5-pin |
DO Sensor InPro6960i/12/120