Styron 683
Measuring cell: DSC821e
Pan: Aluminum standard 40 µl, pierced lid
Sample preparation: Disk cut from pellet, 12 mg
DSC measurement: First heating run from 30 to 250ºC at 10 K/min, quench-cooled to room temperature the sample changer, second heating run from 30 to 250ºC at 10 K/min.
Atmosphere: Nitrogen, 50 cm3/min
The first run has not been evaluated because it is distorted due to the effects of thermal history. The second run shows the results and additional information given by the glass transition evaluation.
The evaluation should be performed on the second heating run.
DSC Curves of PS | Thermal Analysis Handbook No.HB218 | Application published in METTLER TOLEDO TA Application Handbook Thermoplastics