Cond Sensor InPro7108/DN50-VAR-VP

Cond Sensor InPro7108/DN50-VAR-VP

4-Electrode Conductivity Sensor.The InPro 7108/DN50-VAR-VP utilizes 4-electrode technology to expand its measurement range in medium to high conductivity solutions.

Wide Measuring Range

WideRange technology enables an almost universal use of this sensor, from ultrapure water to drinking water and salt solutions.

Withstands Sterilization Cycles

The sensor is suitable for CIP/SIP applications in Pharma and Food & Beverage, as well in other applications such as phase and media separation.

Withstands Aggressive Procedures

This conductivity sensor has been especially designed and ruggedly built to withstand even the most rigorous cleaning procedures.

Material No.: 52002790

Call for Quote

Features

Compact Sensor Design

The conductivity sensor has a very compact design, which allows installation even in small pipes. At the same time, the sensor design also meets the highest hygienic requirements.

Process Reliability and Easy Connection

The IP68 rated VarioPin connector system ensures excellent signal transmission and avoids problems associated with humidity.

Low Maintenance Due to Flat Surface

The smooth flat surface design prevents any significant fouling. This increases the reliability of the sensor, and only low maintenance is required.

Specifications

Specifications - Cond Sensor InPro7108/DN50-VAR-VP

Temperature Sensor
PT 1000
Cell Constant
0.25 cm-1
Cable Connection
Vario Pin (IP 68)
Cell Constant (nominal)
0.25 cm–1
Measurement Technology
Conductive
Process Connection
VAR050
Sterilizable
Yes
Operating Temperature (Fahrenheit)
14...284 °F
Operating Temperature (Celsius)
-10...140 °C
Mechanical Pressure Resistance (PSI)
250 psi (77 °F), 100 psi (203 °F)
Model Family
4-electrode sensor
Hazardous Area Approvals
Yes
Measured Pressure Resistance
0 – 100 psig @ 203 °F
0 – 246 psig @ 77 °F (0 – 7 barg @ 95 °C
0 – 17 barg @ 25 °C)
Electrode Material
Sterilizable, 316 L
Primary Media/Application
Other Water Processes
Pharmaceutical Waters
Chemical Processes
Other Industrial Processes
Pharmaceutical/Life Science Production
Hygienic/Biocompatibility Approvals
No
Mechanical Pressure Resistance (Bar)
17 bar (25 °C), 7 bar (95 °C)
Sensor Material
PEEK (FDA)
System accuracy
± 5 % or better
Measurement Range
0.02-650 mS/cm

Accessories

Consumables

Software

More Information

Get Your Quote
Cond Sensor InPro7108/DN50-VAR-VP
FDA
USP Class VI