Rainin classic single channel pipette

Rainin Classic Pipette

Rugged Construction, Dependable Pipetting Performance

A manual pipette is a handheld instrument used in life science laboratories to transfer small amounts of liquids from various well plates, tubes, flasks and reservoirs. The Rainin Classic manual single-channel pipette is an excellent choice for basic pipetting protocols. Robust and durable, this universal-fit pipette will provide years of reproducible performance with low risk of out-of-tolerance errors. The Rainin Classic is available in the following volume ranges: 0.1 μL – 5000 μL.

Call for Quote
View Results ()
Filter ()

Add 1 or 2 more products to compare

Documents

Rainin Classic Manual Pipette Operating Instructions
This document provides a complete overview of the operating instructions for the Rainin Classic Manual Pipette.
Rainin Classic Manual Pipette Datasheet
Rainin Classic improves traditional pipetting. Robust and durable, it provides years of reproducible performance with low out-of-tolerance errors.
800 472 4646 (800 4 RAININ)
Call Service

Expert Service – Tailored to Fit Your Pipetting Needs

METTLER TOLEDO supports and services your pipettes through their entire life-cycle, from good pipetting techniques to preventive maintenance and calibration to equipment repair.

Uptime
Support & Repair
Performance
Maintenance & Optimization
Compliance
Calibration & Quality
Expertise
Training & Consulting

More Information

Product Description

Eight models of the Rainin Classic continuously adjustable digital pipette cover the entire volume range from 0.1 μL to 10 mL. Rainin Classic pipettes are not limited to fixed volume increments but can be set to any volume in range, e.g. 6.6, 133.3, 377, 2228 μL. The piston stroke is set by adjusting a micrometer coupled directly to a digital volume indicator, which reads in microliters. The digital volume indicator simplifies volume setting and virtually eliminates calculation errors.

All 2 to 5000 μL Rainin Classic models are equipped with a stainless steel piston and an o-ring sealing system, which only needs a small amount of oil to function. The seal in the 10 mL model requires grease for lubrication.

A stainless steel tip ejector is provided with all models (except PR-5000 and PR-10ML) for safe disposal of used tips. The ejector has a quick-release mechanism.

Tip Selection

Rainin Classic pipettes are calibrated using Rainin tips. Performance to published specifications is guaranteed only when Rainin tips are used. Graduation marks on each Rainin tip are useful for quick volume checks.

Rainin pipettes and tips are designed together as a pipetting system. Molded from premium-grade virgin polypropylene, all Rainin tips are BioClean and totally inert, so you can be assured of the best pipetting results.

How to Mount a Universal-Fit Pipette Tip

To mount a universal-fit pipette tip, press the shaft into the end of the tip with light force. The tip will seal properly on the shaft with minimal force — do not use more force than is required.

  • Tips must seal properly on the shaft to assure an air-tight seal and avoid leaks or poor accuracy.
  • Tips must be soft and flexible so that the shaft is not scratched or worn prematurely.
  • Tips must be free from microscopic flash and particulates.
  • The tip orifice must be the correct size, and orifice size and geometry must be consistent from tip to tip. Otherwise, accuracy and precision will be affected.
  • Interior and exterior surfaces must be clear, smooth, and hydrophobic to avoid retention of liquid. Too much retention results in poor accuracy and reproducibility.

 

Genuine Rainin tips easily pass these requirements. Rainin cannot accept responsibility for poor performance resulting from the use of tips by other manufacturers.