The GPro 500 carbon monoxide percent gas analyzer is a unique tunable diode laser spectrometer designed for measurement of CO in a gas stream at high temperatures. It uses a folded-path laser beam design for simple installation and CO% measurement.
Top Performance in CO% Measurement
A CO% analyzer for challenging applications, the GPro 500 provides reliable measurement in FCC units, PTA, steam reforming, and hydrogen production.
Low Maintenance and Operating Costs
This CO% gas analyzer is designed to operate at high temperatures, in situ without a conditioning system, reducing the total cost of ownership.
Easy Installation
The GPro 500 is an alignment-free TDL gas analyzer, meaning that the typical challenges of TDL installation and alignment are significantly reduced.
Gas Measured | Carbon monoxide (CO%) |
Lower Detection Limit | 1500 ppm-v |
Measurement range | 0-100% |
Accuracy | 2% of reading or 1500 ppm, whichever is greater |
Linearity | Better than 1% |
Resolution | 1500 ppm-v |
Drift | Negligible (<2% of measurement range between maintenance intervals) |
Sampling rate | 1 second |
Response time (T90) | CO in N₂ 300 ppm-v to 0% in <4 seconds |
Repeatability | ±0.25% of reading or 0.75-v CO (whichever is greater) |
Process pressure range | 0.8 bar - 1.5 bar (abs)/11.6 psi - 21.7 psi (abs) |
Process temperature range | 0-250 °C (23-482 °F); Optional (for probe installation) 0-600 °C (0-1112 °F) with additional thermal barrier |
Effective path length | 50 mm - 10 m, depending on adaption |
Process Adaption or Sensor | Sensor |