PPA, Amodel, injection-molded parts, normal with 75% inorganic filler
Measuring cell: DSC820 with sample changer
Pan: Aluminum standard 40, pierced lid
Sample preparation: First heating run from 35 to 350 C at 20 K/min. The sample was then quench-cooled to room temperature in the sample changer and a second heating run performed under the same conditions
Atmosphere: Nitrogen, 50 cm3/min
In the injection-molded part as supplied (first measurement), the PPA is semicrystalline and the step height of the DSC curve at the glass transition is relatively small. There is practically no cold crystallization. The quench-cooled samples are amorphous and exhibit a clearer glass transition and cold crystallization.
Analysis of the DSC curves provides information on the thermal history, crystallinity and polymer count.
DSC Measurements of PPA | Thermal Analysis Handbook No.HB241 | Application published in METTLER TOLEDO TA Application Handbook Thermoplastics