This article investigates the quality of temperature and heat flow adjustments in the low temperature range. For the highest demands of accuracy, the use of at least 4 reference substances is recommended. The best results are achieved when the transition temperatures are equally spaced and have the largest possible temperature differences between each other. Cyclopentane as a reference substance is not recommended.
UserCom | | Page | | Keywords |
UserCom 9 | 1 | adjustment, reference substances |
Low-temperature calibration | Thermal Analysis Application No. UC 91 | Application published in METTLER TOLEDO Thermal Analysis UserCom 9